Search

Particle Size Analyzer S3500

Request a quote
Create PDF

Language:

The Microtrac S3500 is a industry-preferred Laser Diffraction (LD) analyzer, ideally suited for various particle characterization tasks. It is the first particle size analyzer that uses three precisely placed red laser diodes to accurately characterize particles like never before.

The patented Tri-Laser System provides accurate, reliable and repeatable particle size analysis for a diverse range of applications by utilizing the proven theory of Mie compensation for spherical particles and the proprietary principle of Modified Mie calculations for non-spherical particles. The S3500 measures particle size from 0.02 to 2,800 µm.

Laser Diffraction with red and blue lasers: BLUEWAVE

Particle Size and Shape Analyzer S3500 Reviews from our customers

"The S3500 has awesome image analysis capability. Using this instrument we have obtained high quality results. "

Manik P. K. Maheswaram

Americhem

"Very easy to use system that provides consistent results. The S3500 has some good features and the customer service is great.  "

Robert Gilmore

Savannah River Nuclear Solutions

"Microtrac gives constant robust results. "

Jean-Francois Grenier

Coalia - Thetford Mines

"Laser-diffraction-based particle size analyzer, very accurate in wet as well dry method of operation, very wide range 0.02-2000 micron. Highly recommended for mining and research industry."

Basant Rath

National Mineral Development Corp.

"Flex software to operate with S3500 is helpful and easy for all operators to set up parameters, use the testing program and get values of data clearly for printing out for saving for the next tests. "

Kiet Nguyen

NQA Equipment & Chemicals

"Instrument was sent in for service - Service/Technical Support Team kept me informed of the progress, and the repair was completed in a timely manner. After-service support was excellent, back up and running with the "new" software same day of delivery."

Justin Lloyd

Technic Inc

Trusted reviews provided by

Typical Applications

Used in various fields such as: beverages, biotechnology, chemicals, food, medicine / pharmaceuticals, metal powders, metals, pigments, geology / metallurgy, ...

chemicals

chemicals

battery materials

battery materials

powders

powders

Technical Data

Measuring range0.02 µm - 2.8 mm
Measuring principleLaser diffraction
Lasers3x Red 780 nm
Laser power3 mW nominal
Detection systemTwo fixed photo-electric detectors with logarithmically spaced segments placed at correct angles for optimal scattered light detection from 0.02 to 165 degrees using 151 detector segments.
DataVolume, number and area distributions as well as percentile and other summary data
Data formatStored in ODBC format in encrypted Microsoft Access Databases to ensure compatibility with external statistical software applications.
Data integrityData integrity may be ensured using FDA 21 CFR Part 11 compliant security features including password protection, electronic signatures and assignable permissions
Measuring time~ 10 to 30 seconds
Power requirementsAC input: 90 - 132 VAC, 47 - 63 Hz, single phase
200 to 265 VAC, 47 - 63 Hz, single phase
Power consumption25 W nominal, 50 W max. (depending on options installed)
Environmental conditionsTemperature: 5° to 40° Celsius (50° to 95° Fahrenheit)
Humidity: 90% RH, non- condensing maximum
Storage temperature: -10° to 50° Celsius (14° to 122° Fahrenheit) (dry only)
Pollution: Degree 2
Physical specificationsCase Material: Steel and impact resistant plastic
Exterior surfaces are finished with corrosion resistant paint or plating
Dimensions (W x H x D)~ 560 x 360 x 460 mm (22 x 14 x 18 in)
Weight~ 27 kg (60 lbs )
Eductor air supply100 psi (689 kPa) maximum pressure
5 CFM (8,5 m3/h) at 50 psi (345 kPa) minimum flow rate
Free of dry contaminants, moisture and oil
VacuumVacuum must exceed 50 CFM

Content may be subject to modifications or corrections